Recall Z-0873-2015— Radiometer America Inc

Class II · Terminated

Class II FDA medical device recall by Radiometer America Inc, initiated 2014-10-20, terminated 2015-10-02. It names one FDA 510(k) clearance: K020935. Product: HemoCue¿ Glucose 201 Microcuvettes, The HemoCue Glucose 201 Microcuvettes are designed for use with the HemoCue Glucose 201Analyzer and the HemoCue Glucose 201 DM Analyzer (hereafter referred to as the HemoCue Glucose 201 Analyzer). HemoCue Glucose 201 Microcuvettes are available in individual packages or in vials. Affected lots identified initially were 110705, 110706, 110715, 110716, 110717, 110718, 110719, and 110720. Subsequent investigation also identified as affected lots 1310227, 1310252, and 1310257. HemoCue¿ Glucose 201 Microcuvettes, The HemoCue Glucose 201 Microcuvettes are designed for use with the HemoCue Glucose 201Analyzer and the HemoCue Glucose 201 DM Analyzer (hereafter referred to as the HemoCue Glucose 201 Analyzer). HemoCue Glucose 201 Microcuvettes are available in individual packages or in vials.. Reason: Batches of HemoCue Glucose 201 Microcuvettes show discoloration and provide results outside the specification at the end of their shelf life, approx. after 6 months. The HemoCue Glucose 201 Analyzer with plasma conversion multiplies the measured whole blood glucose value by a factor of 1.11 and displays a plasma equivalent glucose result. HemoCue Glucose 201 Microcuvettes are for In Vitro Diag.

Recalling firm
Radiometer America Inc
Classification
Class II
Status
Terminated
Initiated
2014-10-20
Terminated
2015-10-02
Firm location
Westlake, OH, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

HemoCue¿ Glucose 201 Microcuvettes, The HemoCue Glucose 201 Microcuvettes are designed for use with the HemoCue Glucose 201Analyzer and the HemoCue Glucose 201 DM Analyzer (hereafter referred to as the HemoCue Glucose 201 Analyzer). HemoCue Glucose 201 Microcuvettes are available in individual packages or in vials. Affected lots identified initially were 110705, 110706, 110715, 110716, 110717, 110718, 110719, and 110720. Subsequent investigation also identified as affected lots 1310227, 1310252, and 1310257. HemoCue¿ Glucose 201 Microcuvettes, The HemoCue Glucose 201 Microcuvettes are designed for use with the HemoCue Glucose 201Analyzer and the HemoCue Glucose 201 DM Analyzer (hereafter referred to as the HemoCue Glucose 201 Analyzer). HemoCue Glucose 201 Microcuvettes are available in individual packages or in vials.

Reason for recall

Batches of HemoCue Glucose 201 Microcuvettes show discoloration and provide results outside the specification at the end of their shelf life, approx. after 6 months. The HemoCue Glucose 201 Analyzer with plasma conversion multiplies the measured whole blood glucose value by a factor of 1.11 and displays a plasma equivalent glucose result. HemoCue Glucose 201 Microcuvettes are for In Vitro Diag