HEMOCUE GLUCOSE 201 SYSTEM, ARTICLE NUMBER 120706— 510(k) K020935
Substantially EquivalentHemocue, Inc.
FDA 510(k) clearance K020935 for HEMOCUE GLUCOSE 201 SYSTEM, ARTICLE NUMBER 120706, Substantially Equivalent on 2002-04-03. Applicant: Hemocue, Inc.. Device class: 2.
Clearance details
- Applicant
- Hemocue, Inc.
- Product code
- Code CGA
- Device class
- Class 2
- Regulation
- 21 CFR 862.1345
- Advisory panel
- Clinical Chemistry
- Date received
- Decision date
- Decision
- Substantially Equivalent
- Clearance type
- Special
- Location
- Potomac, MD, US
Recent devices under product code CGA
view allMore clearances from Hemocue, Inc.
view allAdverse events under product code CGA
product code CGA- Death
- 2
- Malfunction
- 1,816
- Total
- 1,818
Source: openFDA MAUDE database, through . Counts aggregate all devices under this product code, not just this specific clearance. Reports are submissions to FDA, not validated events. Search MAUDE directly →
Recalls naming K020935
K-number listed on FDA's recall record- Z-2070-2025 — The HemoCue Glucose 201 Microcuvettes are designed for use with the HemoCue Glucose 201 (or DM) Analyzer with plasma conversion that multiplies the measured whole blood glucose value by a factor of 1.112 and displays a plasma equivalent glucose result. HemoCue Glucose 201 Microcuvettes are available in individual packages or in vials. The box contains 4 vials with 25 microcuvettes in each vial.
- Z-0873-2015 — HemoCue¿ Glucose 201 Microcuvettes, The HemoCue Glucose 201 Microcuvettes are designed for use with the HemoCue Glucose 201Analyzer and the HemoCue Glucose 201 DM Analyzer (hereafter referred to as the HemoCue Glucose 201 Analyzer). HemoCue Glucose 201 Microcuvettes are available in individual packages or in vials. Affected lots identified initially were 110705, 110706, 110715, 110716, 110717, 110718, 110719, and 110720. Subsequent investigation also identified as affected lots 1310227, 1310252, and 1310257. HemoCue¿ Glucose 201 Microcuvettes, The HemoCue Glucose 201 Microcuvettes are designed for use with the HemoCue Glucose 201Analyzer and the HemoCue Glucose 201 DM Analyzer (hereafter referred to as the HemoCue Glucose 201 Analyzer). HemoCue Glucose 201 Microcuvettes are available in individual packages or in vials.
- Z-2190-2013 — HemoCue¿ Glucose 201 Microcuvettes, 25 Microcuvettes Individually Packaged. The HemoCue Glucose 201 Microcuvettes are designed for use with the HemoCue Glucose 201 Analyzer and the HemoCue Glucose 201 DM Analyzer. Quantitative determination of glucose in whole blood using a specially designed analyzer, the HemoCue Glucose 201 with Plasma conversion. The quantitative determination of the instant blood glucose concentration in circulation supplements the clinical evidence in the diagnosis and treatment of patients with diabetes as well as monitoring of neonatal blood glucose levels.
Matched by K-number from FDA's Recall Enterprise System record (openFDA device/recallk_numbers) — these recalls explicitly list K020935.
Data sourced from openFDA. This site is unofficial and independent of the FDA.