Recall Z-3358-2018— B. Braun Medical, Inc.

Class II · Terminated

Class II FDA medical device recall by B. Braun Medical, Inc., initiated 2018-07-20, terminated 2020-05-13. FDA's recall record names no specific 510(k) clearance for this event. Product: CE18TKDP CONT EPID/DURAPREP, Material Number 332215 Connection device used by physicians to provide various anesthetic and fluid administration devices with a single, common access point to a catheter for delivery of anesthetics. The connector is used in conjunction with the catheter for continuous administration of anesthetic fluids.. Reason: The catheter connectors may not stay closed during in and in some cases, leakage or disconnection of the catheters from the catheter connectors has been observed..

Recalling firm
B. Braun Medical, Inc.
Classification
Class II
Status
Terminated
Initiated
2018-07-20
Terminated
2020-05-13
Firm location
Allentown, PA, United States

Product description

CE18TKDP CONT EPID/DURAPREP, Material Number 332215 Connection device used by physicians to provide various anesthetic and fluid administration devices with a single, common access point to a catheter for delivery of anesthetics. The connector is used in conjunction with the catheter for continuous administration of anesthetic fluids.

Reason for recall

The catheter connectors may not stay closed during in and in some cases, leakage or disconnection of the catheters from the catheter connectors has been observed.