Recall Z-3299-2026— Olympus Corporation of the Americas

Class II · Ongoing

Class II FDA medical device recall by Olympus Corporation of the Americas, initiated 2026-08-24. It names one FDA 510(k) clearance: K890328. Product: Olympus M3 Gold Autoclavable Foroblique Telescope 30. Model Number: M3-30A.. Reason: There are identified discrepancies between the sterilization parameters listed in the Instructions for Use and the sterilization parameters listed in the Japanese package insert..

Recalling firm
Olympus Corporation of the Americas
Classification
Class II
Status
Ongoing
Initiated
2026-08-24
Firm location
Center Valley, PA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Olympus M3 Gold Autoclavable Foroblique Telescope 30. Model Number: M3-30A.

Reason for recall

There are identified discrepancies between the sterilization parameters listed in the Instructions for Use and the sterilization parameters listed in the Japanese package insert.