Recall Z-3263-2018— Biomet UK Ltd.
Class II · Terminated
Class II FDA medical device recall by Biomet UK Ltd., initiated 2018-08-09, terminated 2020-05-12. It names one FDA 510(k) clearance: K080685. Product: HipLOC Hip Fracture Fixation System-Hiploc System 135¿ Plate 38 mm Barrel 4 Hole Plate Item Number: 2110-135-004 Product Usage: The HipLOC Compression Hip Screw may be used primarily for fixation of bone fractures. Specific indications include open reduction and internal fixation of a wide variety of fractures of the proximal femur: intracapsular fractures, intertrochanteric fractures and subtrochanteric fractures. Reason: HipLOC System Plate Barrel due to an undersized bore on the HipLOC plate barrel, can result in the lag screw not fitting through the barrel.
- Recalling firm
- Biomet UK Ltd.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2018-08-09
- Terminated
- 2020-05-12
- Firm location
- Bridgend, United Kingdom
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
HipLOC Hip Fracture Fixation System-Hiploc System 135¿ Plate 38 mm Barrel 4 Hole Plate Item Number: 2110-135-004 Product Usage: The HipLOC Compression Hip Screw may be used primarily for fixation of bone fractures. Specific indications include open reduction and internal fixation of a wide variety of fractures of the proximal femur: intracapsular fractures, intertrochanteric fractures and subtrochanteric fractures
Reason for recall
HipLOC System Plate Barrel due to an undersized bore on the HipLOC plate barrel, can result in the lag screw not fitting through the barrel
Data sourced from openFDA. This site is unofficial and independent of the FDA.