Recall Z-3262-2018— Biomet UK Ltd.

Class II · Terminated

Class II FDA medical device recall by Biomet UK Ltd., initiated 2018-08-09, terminated 2020-05-12. It names one FDA 510(k) clearance: K080685. Product: HipLOC Hip Fracture Fixation System-Hiploc System 135¿ Plate 38 mm Barrel 3 Hole Plate Item Number: 2110-135-003 Product Usage: The HipLOC Compression Hip Screw may be used primarily for fixation of bone fractures. Specific indications include open reduction and internal fixation of a wide variety of fractures of the proximal femur: intracapsular fractures, intertrochanteric fractures and subtrochanteric fractures. Reason: HipLOC System Plate Barrel due to an undersized bore on the HipLOC plate barrel, can result in the lag screw not fitting through the barrel.

Recalling firm
Biomet UK Ltd.
Classification
Class II
Status
Terminated
Initiated
2018-08-09
Terminated
2020-05-12
Firm location
Bridgend, United Kingdom

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

HipLOC Hip Fracture Fixation System-Hiploc System 135¿ Plate 38 mm Barrel 3 Hole Plate Item Number: 2110-135-003 Product Usage: The HipLOC Compression Hip Screw may be used primarily for fixation of bone fractures. Specific indications include open reduction and internal fixation of a wide variety of fractures of the proximal femur: intracapsular fractures, intertrochanteric fractures and subtrochanteric fractures

Reason for recall

HipLOC System Plate Barrel due to an undersized bore on the HipLOC plate barrel, can result in the lag screw not fitting through the barrel

Data sourced from openFDA. This site is unofficial and independent of the FDA.