Recall Z-3257-2026— Baxter Healthcare Corporation

Class II · Ongoing

Class II FDA medical device recall by Baxter Healthcare Corporation, initiated 2026-08-03. It names one FDA 510(k) clearance: K200988. Product: Brand Name: VOLARA System Product Name: VOLARA, HC, NA A Model/Catalog Number: M08594A Software Version: N/A Product Description: VOLARA, HOMECARE, NORTH AMERICA Component: No. Reason: Contamination within the release test equipment flow path could have affected the accuracy of the nebulizer flow rate. Affected devices returned to firm and testing confirmed devices meet specifications..

Recalling firm
Baxter Healthcare Corporation
Classification
Class II
Status
Ongoing
Initiated
2026-08-03
Firm location
Deerfield, IL, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Brand Name: VOLARA System Product Name: VOLARA, HC, NA A Model/Catalog Number: M08594A Software Version: N/A Product Description: VOLARA, HOMECARE, NORTH AMERICA Component: No

Reason for recall

Contamination within the release test equipment flow path could have affected the accuracy of the nebulizer flow rate. Affected devices returned to firm and testing confirmed devices meet specifications.

Data sourced from openFDA. This site is unofficial and independent of the FDA.