Recall Z-3243-2018— Terumo Cardiovascular Systems Corporation
Class II · Terminated
Class II FDA medical device recall by Terumo Cardiovascular Systems Corporation, initiated 2018-08-28, terminated 2020-04-23. It names one FDA 510(k) clearance: K172220. Product: Terumo 220/240V AC, Advanced Perfusion System Platform (APS) for use in cardiopulmonary bypass procedures. Catalog Number 801764. Reason: Screws on a Terumo¿ Advanced Perfusion System 1 Power Manager Board (the base unit circuit board) were not torqued to specification. It is possible that a screw which has not been properly torqued can go undetected during manufacturing testing and may loosen over time, resulting in the potential for an electrical short in the system.
- Recalling firm
- Terumo Cardiovascular Systems Corporation
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2018-08-28
- Terminated
- 2020-04-23
- Firm location
- Ann Arbor, MI, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Terumo 220/240V AC, Advanced Perfusion System Platform (APS) for use in cardiopulmonary bypass procedures. Catalog Number 801764
Reason for recall
Screws on a Terumo¿ Advanced Perfusion System 1 Power Manager Board (the base unit circuit board) were not torqued to specification. It is possible that a screw which has not been properly torqued can go undetected during manufacturing testing and may loosen over time, resulting in the potential for an electrical short in the system