Recall Z-3239-2017— Firefly Medical, Inc.

Class II · Terminated

Class II FDA medical device recall by Firefly Medical, Inc., initiated 2017-08-24, terminated 2019-11-01. FDA's recall record names no specific 510(k) clearance for this event. Product: IVEA Product Part Number 16100 Model 500A. Reason: The dowel pins (P/N 16082) that fit in the left and right leg pivot castings (P/Ns 16049/16056) can rotate and gradually maneuver out of the leg casting after several hundred deployment and folding cycles (designed for easy storage) of the IVEA unit..

Recalling firm
Firefly Medical, Inc.
Classification
Class II
Status
Terminated
Initiated
2017-08-24
Terminated
2019-11-01
Firm location
Fort Collins, CO, United States

Product description

IVEA Product Part Number 16100 Model 500A

Reason for recall

The dowel pins (P/N 16082) that fit in the left and right leg pivot castings (P/Ns 16049/16056) can rotate and gradually maneuver out of the leg casting after several hundred deployment and folding cycles (designed for easy storage) of the IVEA unit.

Data sourced from openFDA. This site is unofficial and independent of the FDA.