Recall Z-3237-2026— Philips North America Llc

Class II · Ongoing

Class II FDA medical device recall by Philips North America Llc, initiated 2026-08-21. It names 4 FDA 510(k) clearances: K251808, K213583, K193215 and 1 more. Product: Ingenia Ambition S and Ingenia Ambition X; Model Numbers (REF): 1. 781359 (S), 2. 782108 (S), 3. 782139 (S), 4. 781356 (X), 5. 782109 (X), 6. 782138 (X);. Reason: A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory..

Recalling firm
Philips North America Llc
Classification
Class II
Status
Ongoing
Initiated
2026-08-21
Firm location
Cambridge, MA, United States

510(k) clearances named in this recall

4 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Ingenia Ambition S and Ingenia Ambition X; Model Numbers (REF): 1. 781359 (S), 2. 782108 (S), 3. 782139 (S), 4. 781356 (X), 5. 782109 (X), 6. 782138 (X);

Reason for recall

A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.

Data sourced from openFDA. This site is unofficial and independent of the FDA.