Recall Z-3233-2017— Sorin Group Italia SRL - CRF

Class II · Terminated

Class II FDA medical device recall by Sorin Group Italia SRL - CRF, initiated 2017-07-24, terminated 2024-01-29. FDA's recall record names no specific 510(k) clearance for this event. Product: Platinium CRT-D 1741, Model Number TDF037U, UDI/GTIN 08031527015484, Implantable cardioverter defibrillator, Biventricular or ventricular antitachycardia pacing, Dual or single chamber arrhythmia detection Product Usage: PLATINIUM VR is indicated for use in patients who are at high risk of sudden cardiac death due to ventricular tachyarrhythmias and who have experienced one of the following situations: - Survival of at least one episode of cardiac arrest (manifested by the loss of consciousness) due to ventricular tachyarrhythmia, - Recurrent, poorly tolerated sustained ventricular tachycardia (VT).. Reason: There is a possibility of overconsumption of certain PLATINIUM Implantable Cardiac Defibrillators (ICDs) and Cardiac Resynchronization Therapy Defibrillators (CRT-Ds) following an ElectroStatic Discharge (ESD) during the implant surgery or a Magnetic Resonance Imaging (MRI) scan, and what actions to take..

Recalling firm
Sorin Group Italia SRL - CRF
Classification
Class II
Status
Terminated
Initiated
2017-07-24
Terminated
2024-01-29
Firm location
Saluggia, VC, N/A, Italy

Product description

Platinium CRT-D 1741, Model Number TDF037U, UDI/GTIN 08031527015484, Implantable cardioverter defibrillator, Biventricular or ventricular antitachycardia pacing, Dual or single chamber arrhythmia detection Product Usage: PLATINIUM VR is indicated for use in patients who are at high risk of sudden cardiac death due to ventricular tachyarrhythmias and who have experienced one of the following situations: - Survival of at least one episode of cardiac arrest (manifested by the loss of consciousness) due to ventricular tachyarrhythmia, - Recurrent, poorly tolerated sustained ventricular tachycardia (VT).

Reason for recall

There is a possibility of overconsumption of certain PLATINIUM Implantable Cardiac Defibrillators (ICDs) and Cardiac Resynchronization Therapy Defibrillators (CRT-Ds) following an ElectroStatic Discharge (ESD) during the implant surgery or a Magnetic Resonance Imaging (MRI) scan, and what actions to take.

Data sourced from openFDA. This site is unofficial and independent of the FDA.