Recall Z-3232-2026— Philips North America Llc
Class II · Ongoing
Class II FDA medical device recall by Philips North America Llc, initiated 2026-08-21. It names 5 FDA 510(k) clearances: K193215, K190461, K063559 and 2 more. Product: Achieva 1.5T, Achieva 1.5T Conversion and Achieva 1.5T Initial system; Model Numbers (REF): 1. 781196 (Achieva 1.5T), 2. 781343 (Achieva 1.5T), 3. 781283 (Achieva 1.5T Conversion), 4. 781346 (Achieva 1.5 Conversion), 5. 781178 (Achieva 1.5T Initial System);. Reason: A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory..
- Recalling firm
- Philips North America Llc
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2026-08-21
- Firm location
- Cambridge, MA, United States
510(k) clearances named in this recall
5 on FDA's record- K193215 — Achieva, Intera, Ingenia, Ingneia CX, Ingenia Elition, and Ingenia Ambition MR Systems
- K190461 — Achieva 1.5T, 3.0T and Intera 1.5T MR Systems
- K063559 — ACHIEVA, INTERA AND PANORAMA 1.0T, RELEASE 2.5
- K052078 — ACHIEVA, INTERA AND PANORAMA 1.0 RELEASE 2-SERIES
- K052013 — ACHIEVA 1.5T & INERA 1.5T FAMILY
K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Achieva 1.5T, Achieva 1.5T Conversion and Achieva 1.5T Initial system; Model Numbers (REF): 1. 781196 (Achieva 1.5T), 2. 781343 (Achieva 1.5T), 3. 781283 (Achieva 1.5T Conversion), 4. 781346 (Achieva 1.5 Conversion), 5. 781178 (Achieva 1.5T Initial System);
Reason for recall
A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.
Data sourced from openFDA. This site is unofficial and independent of the FDA.