Recall Z-3231-2026— Biomerieux Inc

Class II · Ongoing

Class II FDA medical device recall by Biomerieux Inc, initiated 2026-07-24. FDA's recall record names no specific 510(k) clearance for this event. Product: VITEK 2 AST-N461 TEST KIT, GRAM NEGATIVE SUSCEPTIBILITY TESTING - AUTOMATED, Reference Number 424811. Reason: The issue is related to card pouch damage associated with a portion of the lot which was produced on a specific card pouching device, in a specific timeframe, causing a potential risk of false resistant results. False resistant results may lead to the use of less effective antibiotic treatments or the association of multiple antibiotics having negative side effects, which could potentially impact patient s health..

Recalling firm
Biomerieux Inc
Classification
Class II
Status
Ongoing
Initiated
2026-07-24
Firm location
Hazelwood, MO, United States

Product description

VITEK 2 AST-N461 TEST KIT, GRAM NEGATIVE SUSCEPTIBILITY TESTING - AUTOMATED, Reference Number 424811

Reason for recall

The issue is related to card pouch damage associated with a portion of the lot which was produced on a specific card pouching device, in a specific timeframe, causing a potential risk of false resistant results. False resistant results may lead to the use of less effective antibiotic treatments or the association of multiple antibiotics having negative side effects, which could potentially impact patient s health.