Recall Z-3205-2026— Howmedica Osteonics Corp.
Class II · Ongoing
Class II FDA medical device recall by Howmedica Osteonics Corp., initiated 2026-07-30. FDA's recall record names no specific 510(k) clearance for this event. Product: SCORPIO + MOBILE BEARING PCS 78-5-0910 SYSTEM ONE BIPOLAR 40MM 6074-4022 SYSTEM ONE BIPOLAR 43MM 6074-4322 SYSTEM ONE BIPOLAR 44MM 6074-4422 Orthopaedic implant components manufactured from UHMWPE (polyethylene). Reason: Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation..
- Recalling firm
- Howmedica Osteonics Corp.
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2026-07-30
- Firm location
- Mahwah, NJ, United States
Product description
SCORPIO + MOBILE BEARING PCS 78-5-0910 SYSTEM ONE BIPOLAR 40MM 6074-4022 SYSTEM ONE BIPOLAR 43MM 6074-4322 SYSTEM ONE BIPOLAR 44MM 6074-4422 Orthopaedic implant components manufactured from UHMWPE (polyethylene)
Reason for recall
Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.
Data sourced from openFDA. This site is unofficial and independent of the FDA.