Recall Z-3191-2026— Datex-Ohmeda, Inc.

Class II · Ongoing

Class II FDA medical device recall by Datex-Ohmeda, Inc., initiated 2026-07-30. It names one FDA 510(k) clearance: K172045. Product: Aisys CS2 with Et Control, Datex-Ohmeda anesthesia system, Part Number 1011-9000-000. Reason: Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling continued controlled or manual ventilation and initiates audible and visual alarms. When this issue occurs, anesthetic agent delivery will be interrupted and requires clinician intervention. If this is not provided, the patient may experience decreased anesthetic agent effect..

Recalling firm
Datex-Ohmeda, Inc.
Classification
Class II
Status
Ongoing
Initiated
2026-07-30
Firm location
Madison, WI, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Aisys CS2 with Et Control, Datex-Ohmeda anesthesia system, Part Number 1011-9000-000

Reason for recall

Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling continued controlled or manual ventilation and initiates audible and visual alarms. When this issue occurs, anesthetic agent delivery will be interrupted and requires clinician intervention. If this is not provided, the patient may experience decreased anesthetic agent effect.

Data sourced from openFDA. This site is unofficial and independent of the FDA.