Recall Z-3190-2018— Siemens Medical Solutions USA, Inc

Class II · Terminated

Class II FDA medical device recall by Siemens Medical Solutions USA, Inc, initiated 2018-07-29, terminated 2019-05-17. It names one FDA 510(k) clearance: K163286. Product: ARTIS Pheno Interventional Fluoroscopic X-Ray System, Model Number 10849000. Reason: Temperature of the detector may fall out of its designed working range (cool down) due to incorrectly configured temperature control unit. This unintentional system behavior may occur while the unit is powered down for long periods (e.g. at night)..

Recalling firm
Siemens Medical Solutions USA, Inc
Classification
Class II
Status
Terminated
Initiated
2018-07-29
Terminated
2019-05-17
Firm location
Malvern, PA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

ARTIS Pheno Interventional Fluoroscopic X-Ray System, Model Number 10849000

Reason for recall

Temperature of the detector may fall out of its designed working range (cool down) due to incorrectly configured temperature control unit. This unintentional system behavior may occur while the unit is powered down for long periods (e.g. at night).

Data sourced from openFDA. This site is unofficial and independent of the FDA.