Recall Z-3189-2026— Datex-Ohmeda, Inc.
Class II · Ongoing
Class II FDA medical device recall by Datex-Ohmeda, Inc., initiated 2026-07-30. It names one FDA 510(k) clearance: K172045. Product: Aisys, Datex-Ohmeda anesthesia system, Part Number 1011-9050-000. Reason: Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling continued controlled or manual ventilation and initiates audible and visual alarms. When this issue occurs, anesthetic agent delivery will be interrupted and requires clinician intervention. If this is not provided, the patient may experience decreased anesthetic agent effect..
- Recalling firm
- Datex-Ohmeda, Inc.
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2026-07-30
- Firm location
- Madison, WI, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Aisys, Datex-Ohmeda anesthesia system, Part Number 1011-9050-000
Reason for recall
Anesthesia Delivery Systems containing certain electronic gas mixer assemblies can automatically transition to Alternate O2 control. In Alternate O2 control, the system delivers 100% O2 to the breathing system enabling continued controlled or manual ventilation and initiates audible and visual alarms. When this issue occurs, anesthetic agent delivery will be interrupted and requires clinician intervention. If this is not provided, the patient may experience decreased anesthetic agent effect.
Data sourced from openFDA. This site is unofficial and independent of the FDA.