Recall Z-3180-2018— bioMerieux, Inc.

Class II · Terminated

Class II FDA medical device recall by bioMerieux, Inc., initiated 2018-08-13, terminated 2020-11-02. FDA's recall record names no specific 510(k) clearance for this event. Product: VITEK(R) 2 AST-P655, For susceptibility testing of Staphylococcus spp. Enterococcus spp., ad S agalactiae against specified antimicrobials, REF 421913 Product Usage: VITEK¿ 2 is an automated system consisting of instruments, software and reagent cards designed for the identification and antimicrobial susceptibility testing of bacteria and yeast. The VITEK¿ 2 utilizes growth-based biochemical patterns to determine identification. The VITEK¿ 2 provides minimal inhibitory concentration (MIC) results for most organism/drug combinations as well as a category interpretation (S, I, or R) consistent with accepted criteria.. Reason: Routine internal review of Device History Records identified six VITEK(R) 2 AST card lots were manufactured with less drug (gentamicin) than required for the Gentamicin High-Level Resistance (ghlr01n) screen test due to an error when making 1500mL and 2000mL media well preparations (solutions)..

Recalling firm
bioMerieux, Inc.
Classification
Class II
Status
Terminated
Initiated
2018-08-13
Terminated
2020-11-02
Firm location
Durham, NC, United States

Product description

VITEK(R) 2 AST-P655, For susceptibility testing of Staphylococcus spp. Enterococcus spp., ad S agalactiae against specified antimicrobials, REF 421913 Product Usage: VITEK¿ 2 is an automated system consisting of instruments, software and reagent cards designed for the identification and antimicrobial susceptibility testing of bacteria and yeast. The VITEK¿ 2 utilizes growth-based biochemical patterns to determine identification. The VITEK¿ 2 provides minimal inhibitory concentration (MIC) results for most organism/drug combinations as well as a category interpretation (S, I, or R) consistent with accepted criteria.

Reason for recall

Routine internal review of Device History Records identified six VITEK(R) 2 AST card lots were manufactured with less drug (gentamicin) than required for the Gentamicin High-Level Resistance (ghlr01n) screen test due to an error when making 1500mL and 2000mL media well preparations (solutions).

Data sourced from openFDA. This site is unofficial and independent of the FDA.