Recall Z-3176-2024— B Braun Medical Inc

Class II · Ongoing

Class II FDA medical device recall by B Braun Medical Inc, initiated 2024-08-15. It names one FDA 510(k) clearance: K080807. Product: STREAMLINE BLOODLINE LONG VERSION, FMC-Tubing to be used to transfer blood between a patient and a haemodialyser during haemodialysis. Model/Catalog Number: SL-2000M2095L. Reason: Incomplete insertion of the patient connector of the Streamline Airless System Hemodialysis Bloodline to the Locksite may result in the potential for occlusion during prograde rinseback procedures. If the patient connector is not fully inserted into the Locksite, the internal septum may not open fully enough to prevent restrictions to flow. may cause delays in therapy while clinicians are forced to switch to retrograde flow method.

Recalling firm
B Braun Medical Inc
Classification
Class II
Status
Ongoing
Initiated
2024-08-15
Firm location
Bethlehem, PA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

STREAMLINE BLOODLINE LONG VERSION, FMC-Tubing to be used to transfer blood between a patient and a haemodialyser during haemodialysis. Model/Catalog Number: SL-2000M2095L

Reason for recall

Incomplete insertion of the patient connector of the Streamline Airless System Hemodialysis Bloodline to the Locksite may result in the potential for occlusion during prograde rinseback procedures. If the patient connector is not fully inserted into the Locksite, the internal septum may not open fully enough to prevent restrictions to flow. may cause delays in therapy while clinicians are forced to switch to retrograde flow method

Data sourced from openFDA. This site is unofficial and independent of the FDA.