Recall Z-3171-2026— GE Healthcare
Class II · Ongoing
Class II FDA medical device recall by GE Healthcare, initiated 2026-07-31. It names 2 FDA 510(k) clearances: K110875, K043415. Product: Core Services Hub V8, a set of backend services including Cross Enterprise (XE) and Workflow Core (WFC) services that enables enterprise imaging workflows, All versions. Reason: Under certain conditions historical imaging data migrated from a legacy system may become incorrectly associated with a different patient. If this occurs, studies or images may be displayed under the wrong patient record, which could lead to misdiagnosis or a delay in patient care..
- Recalling firm
- GE Healthcare
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2026-07-31
- Firm location
- Chicago, IL, United States
510(k) clearances named in this recall
2 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Core Services Hub V8, a set of backend services including Cross Enterprise (XE) and Workflow Core (WFC) services that enables enterprise imaging workflows, All versions
Reason for recall
Under certain conditions historical imaging data migrated from a legacy system may become incorrectly associated with a different patient. If this occurs, studies or images may be displayed under the wrong patient record, which could lead to misdiagnosis or a delay in patient care.
Data sourced from openFDA. This site is unofficial and independent of the FDA.