Recall Z-3170-2024— Siemens Healthcare Diagnostics, Inc.

Class II · Ongoing

Class II FDA medical device recall by Siemens Healthcare Diagnostics, Inc., initiated 2024-08-15. FDA's recall record names no specific 510(k) clearance for this event. Product: ADVIA Chemistry Microalbumin_2 (¿ALB_2) -For in vitro diagnostic use in the quantitative measurement of microalbumin in human urine on ADVIA Chemistry systems/ Atellica CH and Atellica CI Analyzers.. Reason: Atellica CH Microalbumin_2 (¿ALB_2) and the ADVIA Chemistry Microalbumin_2 (¿ALB_2) are not meeting the High-Dose Hook Effect claim/Prozone Effect claim as stated in the Instructions for Use (IFU) on the Atellica CH, Atellica CI Analyzers, ADVIA 1800 Chemistry Systems, ADVIA 2400 Chemistry Systems, and ADVIA Chemistry XPT Systems, may result in erroneously depressed microalbumin patient results..

Recalling firm
Siemens Healthcare Diagnostics, Inc.
Classification
Class II
Status
Ongoing
Initiated
2024-08-15
Firm location
Tarrytown, NY, United States

Product description

ADVIA Chemistry Microalbumin_2 (¿ALB_2) -For in vitro diagnostic use in the quantitative measurement of microalbumin in human urine on ADVIA Chemistry systems/ Atellica CH and Atellica CI Analyzers.

Reason for recall

Atellica CH Microalbumin_2 (¿ALB_2) and the ADVIA Chemistry Microalbumin_2 (¿ALB_2) are not meeting the High-Dose Hook Effect claim/Prozone Effect claim as stated in the Instructions for Use (IFU) on the Atellica CH, Atellica CI Analyzers, ADVIA 1800 Chemistry Systems, ADVIA 2400 Chemistry Systems, and ADVIA Chemistry XPT Systems, may result in erroneously depressed microalbumin patient results.

Data sourced from openFDA. This site is unofficial and independent of the FDA.