Recall Z-3167-2026— GE Healthcare
Class II · Ongoing
Class II FDA medical device recall by GE Healthcare, initiated 2026-07-31. It names 2 FDA 510(k) clearances: K110875, K043415. Product: Enterprise Archive V8, software product for receiving, archiving, and sending electronic medical data, All versions. Reason: Under certain conditions historical imaging data migrated from a legacy system may become incorrectly associated with a different patient. If this occurs, studies or images may be displayed under the wrong patient record, which could lead to misdiagnosis or a delay in patient care..
- Recalling firm
- GE Healthcare
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2026-07-31
- Firm location
- Chicago, IL, United States
510(k) clearances named in this recall
2 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Enterprise Archive V8, software product for receiving, archiving, and sending electronic medical data, All versions
Reason for recall
Under certain conditions historical imaging data migrated from a legacy system may become incorrectly associated with a different patient. If this occurs, studies or images may be displayed under the wrong patient record, which could lead to misdiagnosis or a delay in patient care.