Recall Z-3152-2017— Zimmer Biomet, Inc.

Class II · Terminated

Class II FDA medical device recall by Zimmer Biomet, Inc., initiated 2017-05-22, terminated 2019-03-29. FDA's recall record names no specific 510(k) clearance for this event. Product: Drill Bit Jacobs Chuck 1.5 mm Diameter 70 mm Length, Item Number/EDI 00480507015, Nonsterile. orthopedic surgical instrument.. Reason: Various trauma and extremity instruments (drill bits and screw taps) were manufactured from raw material with an anomaly. The anomaly may be on or below the surface and can increase the risk of instrument fracture..

Recalling firm
Zimmer Biomet, Inc.
Classification
Class II
Status
Terminated
Initiated
2017-05-22
Terminated
2019-03-29
Firm location
Warsaw, IN, United States

Product description

Drill Bit Jacobs Chuck 1.5 mm Diameter 70 mm Length, Item Number/EDI 00480507015, Nonsterile. orthopedic surgical instrument.

Reason for recall

Various trauma and extremity instruments (drill bits and screw taps) were manufactured from raw material with an anomaly. The anomaly may be on or below the surface and can increase the risk of instrument fracture.