Recall Z-3143-2026— GE Medical Systems Information Technologies Inc

Class II · Ongoing

Class II FDA medical device recall by GE Medical Systems Information Technologies Inc, initiated 2026-07-30. It names one FDA 510(k) clearance: K162012. Product: Carescape Central Station CSCS V3 MAS800 Desktop Part Number 5867474-04, 5867474-13 Product code STU CSCS V3 MAI800 Integrated Part Number 5867474-03 Product Code STV. Reason: Device may not resume patient monitoring for at least 3 minutes following system restart. Issue can lead to potential delay in recognition of clinically significant events..

Recalling firm
GE Medical Systems Information Technologies Inc
Classification
Class II
Status
Ongoing
Initiated
2026-07-30
Firm location
Waukesha, WI, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Carescape Central Station CSCS V3 MAS800 Desktop Part Number 5867474-04, 5867474-13 Product code STU CSCS V3 MAI800 Integrated Part Number 5867474-03 Product Code STV

Reason for recall

Device may not resume patient monitoring for at least 3 minutes following system restart. Issue can lead to potential delay in recognition of clinically significant events.

Data sourced from openFDA. This site is unofficial and independent of the FDA.