Class II · Ongoing

Class II FDA medical device recall by SEER MEDICAL PTY LTD, initiated 2024-08-12. It names one FDA 510(k) clearance: K212788. Product: Seer Home System is an ambulatory electroencephalograph (EEG) system intended for at-home use for extended EEG and video EEG (vEEG) recordings and analysis. The Seer Home System includes the Sight Monitoring Hub, the Sense wearable, Monitoring Portal and associated accessory components (Disposable EEG Wires, WaterTabs, WaterTabs Application Kit, and Take-home Kit).. Reason: During CAPA investigations of sight production inconsistencies, it was highlighted that previous Electromagnetic Compatibility (EMC) assessments related to RF radiated emissions may not be reasonably equivalent to the current product due to design changes in the system. A new CAPA was opened to investigate the radiated emissions compliance issues. - Seer Home Systems distributed and retained in the field were not compliant with Electromagnetic Compatibility (EMC) testing according to IEC 60601-1-2 - Seer Home Systems, including accessories, were manufactured and distributed without application of proper design change and manufacturing controls - Seer Home Systems were distributed with labeling deficiencies.

Recalling firm
SEER MEDICAL PTY LTD
Classification
Class II
Status
Ongoing
Initiated
2024-08-12
Firm location
Melbourne, Australia

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Seer Home System is an ambulatory electroencephalograph (EEG) system intended for at-home use for extended EEG and video EEG (vEEG) recordings and analysis. The Seer Home System includes the Sight Monitoring Hub, the Sense wearable, Monitoring Portal and associated accessory components (Disposable EEG Wires, WaterTabs, WaterTabs Application Kit, and Take-home Kit).

Reason for recall

During CAPA investigations of sight production inconsistencies, it was highlighted that previous Electromagnetic Compatibility (EMC) assessments related to RF radiated emissions may not be reasonably equivalent to the current product due to design changes in the system. A new CAPA was opened to investigate the radiated emissions compliance issues. - Seer Home Systems distributed and retained in the field were not compliant with Electromagnetic Compatibility (EMC) testing according to IEC 60601-1-2 - Seer Home Systems, including accessories, were manufactured and distributed without application of proper design change and manufacturing controls - Seer Home Systems were distributed with labeling deficiencies