Recall Z-3122-2024— BioFire Diagnostics, LLC

Class II · Ongoing

Class II FDA medical device recall by BioFire Diagnostics, LLC, initiated 2024-08-07. It names one FDA 510(k) clearance: K143178. Product: FILMARRAY 2.0 INSTRUMENT, REF: FLM2-ASY-0001, FLM2-ASY-0001-W. Reason: Some automated in vitro diagnostic modules and instruments manufactured or serviced between May 2022 and May 2024 were calibrated with internal equipment that may lead modules and instruments to report higher melting temperatures, which may lead to erroneous results including false positive Salmonella, Giardia lamblia, false positive/negative Norovirus, and false negative MRSA..

Recalling firm
BioFire Diagnostics, LLC
Classification
Class II
Status
Ongoing
Initiated
2024-08-07
Firm location
Salt Lake City, UT, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

FILMARRAY 2.0 INSTRUMENT, REF: FLM2-ASY-0001, FLM2-ASY-0001-W

Reason for recall

Some automated in vitro diagnostic modules and instruments manufactured or serviced between May 2022 and May 2024 were calibrated with internal equipment that may lead modules and instruments to report higher melting temperatures, which may lead to erroneous results including false positive Salmonella, Giardia lamblia, false positive/negative Norovirus, and false negative MRSA.

Data sourced from openFDA. This site is unofficial and independent of the FDA.