Recall Z-3118-2017— Beckman Coulter Inc.
Class II · Terminated
Class II FDA medical device recall by Beckman Coulter Inc., initiated 2017-07-20, terminated 2018-01-04. It names one FDA 510(k) clearance: K081938. Product: Sekisui Diagnostics Sekure Chemistry Acetaminophen L3K , Ref. 506-30. The firm names on the label are Sekisui Diagnostics (UK) Limited, United Kingdom, and Sekisui Diagnostics P.E.I. Inc., Canada. Used in laboratory testing.. Reason: Reagent carryover from Sekisui ACTH when utilized on the BEC AU 680 analyzer may lead to falsely elevated BEC Lipase recovery exceeding the total precision specification..
- Recalling firm
- Beckman Coulter Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2017-07-20
- Terminated
- 2018-01-04
- Firm location
- Brea, CA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Sekisui Diagnostics Sekure Chemistry Acetaminophen L3K , Ref. 506-30. The firm names on the label are Sekisui Diagnostics (UK) Limited, United Kingdom, and Sekisui Diagnostics P.E.I. Inc., Canada. Used in laboratory testing.
Reason for recall
Reagent carryover from Sekisui ACTH when utilized on the BEC AU 680 analyzer may lead to falsely elevated BEC Lipase recovery exceeding the total precision specification.
Data sourced from openFDA. This site is unofficial and independent of the FDA.