Recall Z-3104-2024— Siemens Healthcare Diagnostics, Inc.

Class II · Ongoing

Class II FDA medical device recall by Siemens Healthcare Diagnostics, Inc., initiated 2024-07-26. It names one FDA 510(k) clearance: K031393. Product: ADVIA Centaur CA 19-9 (250 Test)-In vitro diagnostic use in the quantitative measurement of the CA 19-9 tumor-associated antigen in human serum using the Atellica CI and Atellica IM Analyzers SMN: 10491244. Reason: Positive bias for some samples around the CA 19-9 upper limit of normal of 35 U/m as compared to the previous lot. The sample bias observed was greatest in the Asia Pacific sample population. The observed bias is not proportional across the assay measuring interval and is less pronounced for sample results above a concentration of 110 U/mL..

Recalling firm
Siemens Healthcare Diagnostics, Inc.
Classification
Class II
Status
Ongoing
Initiated
2024-07-26
Firm location
East Walpole, MA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

ADVIA Centaur CA 19-9 (250 Test)-In vitro diagnostic use in the quantitative measurement of the CA 19-9 tumor-associated antigen in human serum using the Atellica CI and Atellica IM Analyzers SMN: 10491244

Reason for recall

Positive bias for some samples around the CA 19-9 upper limit of normal of 35 U/m as compared to the previous lot. The sample bias observed was greatest in the Asia Pacific sample population. The observed bias is not proportional across the assay measuring interval and is less pronounced for sample results above a concentration of 110 U/mL.