Recall Z-3102-2017— Encore Medical, Lp

Class II · Terminated

Class II FDA medical device recall by Encore Medical, Lp, initiated 2017-08-04, terminated 2023-06-28. FDA's recall record names no specific 510(k) clearance for this event. Product: MODIFIED/SPECIAL INSTRUMENT, TURON GLENOID REAMER, REAM AND RUN, Catalog Numbers: REF S-200956 - 5mm PIN, SIZE 46 REF S-200957 - 5mm PIN, SIZE 50 REF S-200958 - 5mm PIN, SIZE 58. Reason: It was reported that the models used to manufacture S-200956/958 had the wrong dimension used to create the spherical radius cutting surface. Device design allows a larger spherical radius cutting surface than intended..

Recalling firm
Encore Medical, Lp
Classification
Class II
Status
Terminated
Initiated
2017-08-04
Terminated
2023-06-28
Firm location
Austin, TX, United States

Product description

MODIFIED/SPECIAL INSTRUMENT, TURON GLENOID REAMER, REAM AND RUN, Catalog Numbers: REF S-200956 - 5mm PIN, SIZE 46 REF S-200957 - 5mm PIN, SIZE 50 REF S-200958 - 5mm PIN, SIZE 58

Reason for recall

It was reported that the models used to manufacture S-200956/958 had the wrong dimension used to create the spherical radius cutting surface. Device design allows a larger spherical radius cutting surface than intended.