Recall Z-3076-2017— Biomerieux Inc
Class II · Terminated
Class II FDA medical device recall by Biomerieux Inc, initiated 2017-07-12, terminated 2020-12-16. It names one FDA 510(k) clearance: K050002. Product: VITEK¿ 2 Compact 30, REF numbers: 27530 and 27530R (clinical), and 27630 and 17630R (industry) software and reagent cards designed for the identification (ID) and antimicrobial susceptibility testing (AST) of bacteria and yeast.. Reason: Customers have reported that some VITEK¿ 2 cards are staying in preliminary status, not finalizing after ejection from the instrument, and not allowing cards in subsequent carousel slots to be processed. The issue was reported to occur on VITEK¿ 2 Compact 15 and Compact 30 systems following a system software update to version 8.01..
- Recalling firm
- Biomerieux Inc
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2017-07-12
- Terminated
- 2020-12-16
- Firm location
- Hazelwood, MO, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
VITEK¿ 2 Compact 30, REF numbers: 27530 and 27530R (clinical), and 27630 and 17630R (industry) software and reagent cards designed for the identification (ID) and antimicrobial susceptibility testing (AST) of bacteria and yeast.
Reason for recall
Customers have reported that some VITEK¿ 2 cards are staying in preliminary status, not finalizing after ejection from the instrument, and not allowing cards in subsequent carousel slots to be processed. The issue was reported to occur on VITEK¿ 2 Compact 15 and Compact 30 systems following a system software update to version 8.01.
Data sourced from openFDA. This site is unofficial and independent of the FDA.