Recall Z-3044-2018— Ortho-Clinical Diagnostics
Class III · Terminated
Class III FDA medical device recall by Ortho-Clinical Diagnostics, initiated 2018-07-11, terminated 2020-08-11. It names one FDA 510(k) clearance: K071801. Product: VITROS Chemistry Products Cl- Slides, Product Code 684 4471 Product Usage For in vitro diagnostic use only. VITROS Chemistry Products Cl- Slides quantitatively measure chloride (Cl-) concentration in serum, plasma and urine using VITROS 250/350/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System. Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis. Product Usage:. Reason: The slides do not meet the maximum allowable interference (MAI) claims for triglyceride interference when using serum samples. Triglyceride interference on the Cl- Slide was observed around a triglyceride concentration 600mg/dL, however the current triglyceride interference claim states interference should not occur below 800mg/dL of triglyceride..
- Recalling firm
- Ortho-Clinical Diagnostics
- Classification
- Class III
- Status
- Terminated
- Initiated
- 2018-07-11
- Terminated
- 2020-08-11
- Firm location
- Rochester, NY, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
VITROS Chemistry Products Cl- Slides, Product Code 684 4471 Product Usage For in vitro diagnostic use only. VITROS Chemistry Products Cl- Slides quantitatively measure chloride (Cl-) concentration in serum, plasma and urine using VITROS 250/350/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System. Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis. Product Usage:
Reason for recall
The slides do not meet the maximum allowable interference (MAI) claims for triglyceride interference when using serum samples. Triglyceride interference on the Cl- Slide was observed around a triglyceride concentration 600mg/dL, however the current triglyceride interference claim states interference should not occur below 800mg/dL of triglyceride.