Recall Z-3039-2020— Ameditech Inc

Class II · Terminated

Class II FDA medical device recall by Ameditech Inc, initiated 2020-08-18, terminated 2024-09-30. It names one FDA 510(k) clearance: K122633. Product: Noble 1 Step+ Cup (AMT Item Number/ AMT Description): NBCA-13A-B /"Noble 1 Step + Cup" THC50/COC150; OPI300/MET500; AMP500/BZO300; BAR300/MTD300; BUPG10/TCA1000; MDMA500/OXY100; w/ Cr, Ni, pH, Bl, S.G. Finished Devices", NBCA-14A-PP/"Noble 1 Step + Cup" (THC50/COC150)+(OPI300/MET500)+ (AMP500/BZO300)+(BAR300/MTD300)+ (BOPG10/TCA1000)+(MDMA500/OXY100)+ (PCP25/PPX300) w/ Cr, Ni, pH, Bl, S.G.". Reason: Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling..

Recalling firm
Ameditech Inc
Classification
Class II
Status
Terminated
Initiated
2020-08-18
Terminated
2024-09-30
Firm location
San Diego, CA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Noble 1 Step+ Cup (AMT Item Number/ AMT Description): NBCA-13A-B /"Noble 1 Step + Cup" THC50/COC150; OPI300/MET500; AMP500/BZO300; BAR300/MTD300; BUPG10/TCA1000; MDMA500/OXY100; w/ Cr, Ni, pH, Bl, S.G. Finished Devices", NBCA-14A-PP/"Noble 1 Step + Cup" (THC50/COC150)+(OPI300/MET500)+ (AMP500/BZO300)+(BAR300/MTD300)+ (BOPG10/TCA1000)+(MDMA500/OXY100)+ (PCP25/PPX300) w/ Cr, Ni, pH, Bl, S.G."

Reason for recall

Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.

Data sourced from openFDA. This site is unofficial and independent of the FDA.