Recall Z-3035-2020— Ameditech Inc
Class II · Terminated
Class II FDA medical device recall by Ameditech Inc, initiated 2020-08-18, terminated 2024-09-30. It names one FDA 510(k) clearance: K122633. Product: Henry Schein One Step + DOA Cup (AMT Item Number/ AMT Description): 570-1066 / "Henry Schein OneStep+ DOA Cup 14 Panel AMP500/BAR/BUPG/BZO/COC150/MDMA/MET500/MOP/MTD/OXY/PCP/PPX/TCA/THC (Bl, Cr, Ni, pH, SG)",. Reason: Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling..
- Recalling firm
- Ameditech Inc
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2020-08-18
- Terminated
- 2024-09-30
- Firm location
- San Diego, CA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Henry Schein One Step + DOA Cup (AMT Item Number/ AMT Description): 570-1066 / "Henry Schein OneStep+ DOA Cup 14 Panel AMP500/BAR/BUPG/BZO/COC150/MDMA/MET500/MOP/MTD/OXY/PCP/PPX/TCA/THC (Bl, Cr, Ni, pH, SG)",
Reason for recall
Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.
Data sourced from openFDA. This site is unofficial and independent of the FDA.