Recall Z-3033-2017— Cenorin, LLC
Class II · Terminated
Class II FDA medical device recall by Cenorin, LLC, initiated 2017-06-07, terminated 2018-05-02. FDA's recall record names no specific 510(k) clearance for this event. Product: ThermaSure(TM), a CENORIN(TM) product, SERIES 300 MEDICAL DEVICE DRYER: (a) Model 330; (b) Model 340 Forced air dryer for medical devices. Reason: Dryer unit not heating. Manufacturing issue causes the terminal block to overheat, char and/or burn..
- Recalling firm
- Cenorin, LLC
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2017-06-07
- Terminated
- 2018-05-02
- Firm location
- Kent, WA, United States
Product description
ThermaSure(TM), a CENORIN(TM) product, SERIES 300 MEDICAL DEVICE DRYER: (a) Model 330; (b) Model 340 Forced air dryer for medical devices
Reason for recall
Dryer unit not heating. Manufacturing issue causes the terminal block to overheat, char and/or burn.
Data sourced from openFDA. This site is unofficial and independent of the FDA.