Recall Z-3025-2017— GE Medical Systems, LLC

Class II · Terminated

Class II FDA medical device recall by GE Medical Systems, LLC, initiated 2017-05-11, terminated 2020-09-09. It names 5 FDA 510(k) clearances: K160618, K142361, K142085 and 2 more. Product: Discovery MR750 3.0T Product Usage: Intended use: The Discovery MR750 3.0Tsystem is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. Depending on the region of interest being imaged, contrast agents may be used. The images produced by the Discovery MR750 3.0T system reflect the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.. Reason: Possible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard..

Recalling firm
GE Medical Systems, LLC
Classification
Class II
Status
Terminated
Initiated
2017-05-11
Terminated
2020-09-09
Firm location
Waukesha, WI, United States

510(k) clearances named in this recall

5 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Discovery MR750 3.0T Product Usage: Intended use: The Discovery MR750 3.0Tsystem is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. Depending on the region of interest being imaged, contrast agents may be used. The images produced by the Discovery MR750 3.0T system reflect the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.

Reason for recall

Possible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard.

Data sourced from openFDA. This site is unofficial and independent of the FDA.