Recall Z-3006-2020— Beckman Coulter Inc.
Class II · Terminated
Class II FDA medical device recall by Beckman Coulter Inc., initiated 2020-08-26, terminated 2024-05-24. FDA's recall record names no specific 510(k) clearance for this event. Product: BECKMAN COULTER ACCESS Immunoassay Systems - Access Unconjugated Estriol, REF 33570.. Reason: Due to alkaline phosphatase (ALP) associated interference causing falsely elevated uE3 (Unconjugated Estriol assay) results..
- Recalling firm
- Beckman Coulter Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2020-08-26
- Terminated
- 2024-05-24
- Firm location
- Brea, CA, United States
Product description
BECKMAN COULTER ACCESS Immunoassay Systems - Access Unconjugated Estriol, REF 33570.
Reason for recall
Due to alkaline phosphatase (ALP) associated interference causing falsely elevated uE3 (Unconjugated Estriol assay) results.