Recall Z-2992-2026— Medline Industries, LP
Class II · Ongoing
Class II FDA medical device recall by Medline Industries, LP, initiated 2026-07-13. It names one FDA 510(k) clearance: K903083. Product: Medline breathing circuits labeled as: 1) HUDSON RCI, CIRCUIT, HTD, DL, NEONATE, 48, Medline Product Number/SKU HUD78007 (Teleflex Product Number/SKU 780-07); 2) HUDSON RCI, DBM-CIRCUIT,VENT,NEONATAL,21 V, Medline Product Number/SKU HUD78009 (Teleflex Product Number/SKU 780-09); 3) HUDSON RCI, DBD-CIRCUIT,VENT,INSIRATORY,HEATED,NEO,4, Medline Product Number/SKU 780-13 (Teleflex Product Number/SKU 780-13); 4) HUDSON RCI, DBD-CIRCUIT,VENT,INSIRATORY,HEATED,NEO,4, Medline Product Number/SKU 780-14 (Teleflex Product Number/SKU 780-14); 5) HUDSON RCI, DBD-CIRCUIT, HTD, DUAL LIMB, NEONATE, Medline Product Number/SKU 780-15 (Teleflex Product Number/SKU 780-15); 6) HUDSON RCI, CIRCUIT,NEO,DUAL,WIRE,60 IN, Medline Product Number/SKU HUD78018) (Teleflex Product Number/SKU 780-18).. Reason: Neonatal and infant heated wire breathing circuits may experience sporadic failure. The heated wire connector (pigtail) at the humidifier may exhibit discoloration, melting, smoke, burning odor, or other signs of localized thermal damage. Device failure may predispose infant to bronchospasm, inspissated secretions, and airway obstruction, as well as inadvertent positive end-expiratory pressure (PEEP), circuit obstruction, or accidental lavage of pooled fluid into the airway, elevated oxygen consumption, and worsening respiratory distress..
- Recalling firm
- Medline Industries, LP
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2026-07-13
- Firm location
- Northfield, IL, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Medline breathing circuits labeled as: 1) HUDSON RCI, CIRCUIT, HTD, DL, NEONATE, 48, Medline Product Number/SKU HUD78007 (Teleflex Product Number/SKU 780-07); 2) HUDSON RCI, DBM-CIRCUIT,VENT,NEONATAL,21 V, Medline Product Number/SKU HUD78009 (Teleflex Product Number/SKU 780-09); 3) HUDSON RCI, DBD-CIRCUIT,VENT,INSIRATORY,HEATED,NEO,4, Medline Product Number/SKU 780-13 (Teleflex Product Number/SKU 780-13); 4) HUDSON RCI, DBD-CIRCUIT,VENT,INSIRATORY,HEATED,NEO,4, Medline Product Number/SKU 780-14 (Teleflex Product Number/SKU 780-14); 5) HUDSON RCI, DBD-CIRCUIT, HTD, DUAL LIMB, NEONATE, Medline Product Number/SKU 780-15 (Teleflex Product Number/SKU 780-15); 6) HUDSON RCI, CIRCUIT,NEO,DUAL,WIRE,60 IN, Medline Product Number/SKU HUD78018) (Teleflex Product Number/SKU 780-18).
Reason for recall
Neonatal and infant heated wire breathing circuits may experience sporadic failure. The heated wire connector (pigtail) at the humidifier may exhibit discoloration, melting, smoke, burning odor, or other signs of localized thermal damage. Device failure may predispose infant to bronchospasm, inspissated secretions, and airway obstruction, as well as inadvertent positive end-expiratory pressure (PEEP), circuit obstruction, or accidental lavage of pooled fluid into the airway, elevated oxygen consumption, and worsening respiratory distress.
Data sourced from openFDA. This site is unofficial and independent of the FDA.