Recall Z-2987-2026— restor3d Inc.

Class II · Ongoing

Class II FDA medical device recall by restor3d Inc., initiated 2026-06-30. It names one FDA 510(k) clearance: K243643. Product: Brand Name: Veritas Standard Reverse Total Shoulder Product Name: Veritas Standard Glenoid Baseplate Model/Catalog Number: 1500-1030; 1500-1035 Software Version: N/A - off-the-shelf hardware product Product Description: The restor3d Reverse Total Shoulder Arthroplasty System is intended for patients requiring a reverse shoulder replacement for patients with a functional deltoid muscle and with a deficient rotator cuff. The restor3d Reverse Total Shoulder Arthroplasty System consists of the glenoid baseplate, glenosphere intraoperatively affixed to the baseplate, humeral stem, and polymer bearing component affixed to the humeral stem. Additionally, the system includes supporting standard instrument trays and all required accessories (e.g., security screws, peripheral screws). The restor3d rTSA System is indicated for use in patients whose shoulder joint has a grossly deficient rotator cuff with severe arthropathy and/or previously failed shoulder joint replacement with a grossly deficient rotator cuff. The Veritas standard glenoid baseplate is the only component within scope of this field action. The patient-specific baseplate is unaffected by this defect and therefore not in scope. Component: Reverse Total Shoulder. Reason: Due to incorrect machining of peripheral screw hole resulting in potential for improper engagement of the peripheral screws..

Recalling firm
restor3d Inc.
Classification
Class II
Status
Ongoing
Initiated
2026-06-30
Firm location
Durham, NC, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Brand Name: Veritas Standard Reverse Total Shoulder Product Name: Veritas Standard Glenoid Baseplate Model/Catalog Number: 1500-1030; 1500-1035 Software Version: N/A - off-the-shelf hardware product Product Description: The restor3d Reverse Total Shoulder Arthroplasty System is intended for patients requiring a reverse shoulder replacement for patients with a functional deltoid muscle and with a deficient rotator cuff. The restor3d Reverse Total Shoulder Arthroplasty System consists of the glenoid baseplate, glenosphere intraoperatively affixed to the baseplate, humeral stem, and polymer bearing component affixed to the humeral stem. Additionally, the system includes supporting standard instrument trays and all required accessories (e.g., security screws, peripheral screws). The restor3d rTSA System is indicated for use in patients whose shoulder joint has a grossly deficient rotator cuff with severe arthropathy and/or previously failed shoulder joint replacement with a grossly deficient rotator cuff. The Veritas standard glenoid baseplate is the only component within scope of this field action. The patient-specific baseplate is unaffected by this defect and therefore not in scope. Component: Reverse Total Shoulder

Reason for recall

Due to incorrect machining of peripheral screw hole resulting in potential for improper engagement of the peripheral screws.