Recall Z-2976-2024— PACIFIC LASERTECH, LLC

Class II · Ongoing

Class II FDA medical device recall by PACIFIC LASERTECH, LLC, initiated 2024-03-01. FDA's recall record names no specific 510(k) clearance for this event. Product: 25-LHP-928, HeNe Laser System, 25 mW. Reason: These laser products fail to comply with 21 CFR 1040.10(f)(4) Key control which requires, Each laser system classified as a Class IIIb or IV laser product shall incorporate a key-actuated master control. The key shall be removable and the laser shall not be operable when the key is removed. , because while the laser is switched on, the key can be removed from the key switch. The correction is replacement with a key switch that captures the key when in the on position..

Recalling firm
PACIFIC LASERTECH, LLC
Classification
Class II
Status
Ongoing
Initiated
2024-03-01
Firm location
San Marcos, CA, United States

Product description

25-LHP-928, HeNe Laser System, 25 mW

Reason for recall

These laser products fail to comply with 21 CFR 1040.10(f)(4) Key control which requires, Each laser system classified as a Class IIIb or IV laser product shall incorporate a key-actuated master control. The key shall be removable and the laser shall not be operable when the key is removed. , because while the laser is switched on, the key can be removed from the key switch. The correction is replacement with a key switch that captures the key when in the on position.

Data sourced from openFDA. This site is unofficial and independent of the FDA.