Recall Z-2973-2026— Beckman Coulter Inc.
Class II · Ongoing
Class II FDA medical device recall by Beckman Coulter Inc., initiated 2026-07-15. It names one FDA 510(k) clearance: K242190. Product: Beckman Coulter DxC 500i Clinical Analyzer Modules using software version 1.5 or lower: 1) DxC 500 AU Module w/ISE, DxC 500i, Part Number C63522; 2) DxC 500 AU Module, DxC 500i, Part Number C63521; 3) Access 2 Module, DxC 500i, Part Number C13252; DxC 500 AU Clinical Chemistry Analyzers. Reason: a software defect in which the Parameter Package (PP) import logic overwrites all predefined calculated test rules, including those previously modified by the user, upon import of a new Parameter Package version. As a result, usercustomized calculated test rules revert silently to their default expressions, potentially generating erroneous patient test results without an explicit error notification. The issue affects instruments running software version 1.5 or lower and only occurs when a new Parameter Package is imported to a system that already has an existing Parameter Package onboard; new installations are not affected..
- Recalling firm
- Beckman Coulter Inc.
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2026-07-15
- Firm location
- Brea, CA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Beckman Coulter DxC 500i Clinical Analyzer Modules using software version 1.5 or lower: 1) DxC 500 AU Module w/ISE, DxC 500i, Part Number C63522; 2) DxC 500 AU Module, DxC 500i, Part Number C63521; 3) Access 2 Module, DxC 500i, Part Number C13252; DxC 500 AU Clinical Chemistry Analyzers
Reason for recall
a software defect in which the Parameter Package (PP) import logic overwrites all predefined calculated test rules, including those previously modified by the user, upon import of a new Parameter Package version. As a result, usercustomized calculated test rules revert silently to their default expressions, potentially generating erroneous patient test results without an explicit error notification. The issue affects instruments running software version 1.5 or lower and only occurs when a new Parameter Package is imported to a system that already has an existing Parameter Package onboard; new installations are not affected.