Recall Z-2967-2020— Tosoh Bioscience Inc

Class II · Terminated

Class II FDA medical device recall by Tosoh Bioscience Inc, initiated 2020-08-07, terminated 2021-06-21. FDA's recall record names no specific 510(k) clearance for this event. Product: Tosoh Bioscience AIA- PACK HCG Calibrator Set-In Vitro Diagnostic for BHCG Part number 020361 The AIA-Pack BHCG Calibrator Set is intended for IN VITRO DIAGNOSTIC USE ONLY for the calibration of the ST AIA-PACK BHCG assay. Reason: Decreased stability for five (5) lots of Calibrator Sets and Calibration Verification Test Sets, resulting in a decrease measured concentration of HCG up to 20%..

Recalling firm
Tosoh Bioscience Inc
Classification
Class II
Status
Terminated
Initiated
2020-08-07
Terminated
2021-06-21
Firm location
Grove City, OH, United States

Product description

Tosoh Bioscience AIA- PACK HCG Calibrator Set-In Vitro Diagnostic for BHCG Part number 020361 The AIA-Pack BHCG Calibrator Set is intended for IN VITRO DIAGNOSTIC USE ONLY for the calibration of the ST AIA-PACK BHCG assay

Reason for recall

Decreased stability for five (5) lots of Calibrator Sets and Calibration Verification Test Sets, resulting in a decrease measured concentration of HCG up to 20%.