Recall Z-2959-2026— Tobii Dynavox Llc

Class II · Ongoing

Class II FDA medical device recall by Tobii Dynavox Llc, initiated 2026-07-08. FDA's recall record names no specific 510(k) clearance for this event. Product: TD Navio Midi Speech Generating Device Model/Catalog Number: 1000798. Reason: Under certain conditions, the battery mounting tabs may become damaged, allowing the battery to move within the device enclosure, potentially leading to an internal short circuit. This may cause localized heating and gas venting within the device enclosure..

Recalling firm
Tobii Dynavox Llc
Classification
Class II
Status
Ongoing
Initiated
2026-07-08
Firm location
Pittsburgh, PA, United States

Product description

TD Navio Midi Speech Generating Device Model/Catalog Number: 1000798

Reason for recall

Under certain conditions, the battery mounting tabs may become damaged, allowing the battery to move within the device enclosure, potentially leading to an internal short circuit. This may cause localized heating and gas venting within the device enclosure.

Data sourced from openFDA. This site is unofficial and independent of the FDA.