Recall Z-2957-2024— Abiomed, Inc.

Class II · Ongoing

Class II FDA medical device recall by Abiomed, Inc., initiated 2024-08-05. FDA's recall record names no specific 510(k) clearance for this event. Product: Impella CP Pump 371 Set, Clinical Trial; Product Code: 0048-0040;. Reason: A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction..

Recalling firm
Abiomed, Inc.
Classification
Class II
Status
Ongoing
Initiated
2024-08-05
Firm location
Danvers, MA, United States

Product description

Impella CP Pump 371 Set, Clinical Trial; Product Code: 0048-0040;

Reason for recall

A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.

Data sourced from openFDA. This site is unofficial and independent of the FDA.