Recall Z-2953-2024— PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

Class II · Ongoing

Class II FDA medical device recall by PHILIPS MEDICAL SYSTEMS NEDERLAND B.V., initiated 2024-08-06. It names one FDA 510(k) clearance: K200917. Product: Philips Azurion 7M20 systems with FlexArm ceiling-mounted system, Interventional Fluoroscopic X-Ray System, Model Numbers 722079 and 722224. Reason: The motorized longitudinal movement of the FlexArm stand may be inconsistent (not smooth) and eventually may become unavailable. This issue is due to grease leakage from the bearings and/or anti-corrosion oil applied on the bearings, because of which the lubrication on the ceiling rail may be excessive, reducing the friction between the rails and the friction wheel..

Recalling firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Classification
Class II
Status
Ongoing
Initiated
2024-08-06
Firm location
Best, Netherlands

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Philips Azurion 7M20 systems with FlexArm ceiling-mounted system, Interventional Fluoroscopic X-Ray System, Model Numbers 722079 and 722224

Reason for recall

The motorized longitudinal movement of the FlexArm stand may be inconsistent (not smooth) and eventually may become unavailable. This issue is due to grease leakage from the bearings and/or anti-corrosion oil applied on the bearings, because of which the lubrication on the ceiling rail may be excessive, reducing the friction between the rails and the friction wheel.

Data sourced from openFDA. This site is unofficial and independent of the FDA.