Recall Z-2952-2020— LONGHORN VACCINES AND DIAGNOSTIC
Class II · Ongoing
Class II FDA medical device recall by LONGHORN VACCINES AND DIAGNOSTIC, initiated 2020-06-03. FDA's recall record names no specific 510(k) clearance for this event. Product: PrimeStore Molecular Transport Medium (PS-MTM), Model Numbers LH-1-2, LH-1-3, LH-1-4, LH-1-5, LH-1-5CC, LH-1-15CC - Product Usage: intended for stabilization, transportation, and inactivation of infectious unprocessed nasal washes, nasopharyngeal, oral/throat swabs* suspected of containing Influenza A virus RNA. PS-MTM is also intended for the stabilization, transportation, and inactivation of infectious unprocessed sputum samples suspected of containing Mycobacterium tuberculosis (MTB) DNA from human samples.. Reason: The tubes may have been delivered to some users without labels. Users may not be able to identify the tubes without these labels. If samples stored or transported in tubes are tested on certain systems or come into contact with bleach they can produce cyanide gas. The solution in these tubes contains guanidine thiocynate and may put laboratory personnel at risk of exposure to cyanide gas when it comes into contact with bleach or reagents containing bleach..
- Recalling firm
- LONGHORN VACCINES AND DIAGNOSTIC
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2020-06-03
- Firm location
- Bethesda, MD, United States
Product description
PrimeStore Molecular Transport Medium (PS-MTM), Model Numbers LH-1-2, LH-1-3, LH-1-4, LH-1-5, LH-1-5CC, LH-1-15CC - Product Usage: intended for stabilization, transportation, and inactivation of infectious unprocessed nasal washes, nasopharyngeal, oral/throat swabs* suspected of containing Influenza A virus RNA. PS-MTM is also intended for the stabilization, transportation, and inactivation of infectious unprocessed sputum samples suspected of containing Mycobacterium tuberculosis (MTB) DNA from human samples.
Reason for recall
The tubes may have been delivered to some users without labels. Users may not be able to identify the tubes without these labels. If samples stored or transported in tubes are tested on certain systems or come into contact with bleach they can produce cyanide gas. The solution in these tubes contains guanidine thiocynate and may put laboratory personnel at risk of exposure to cyanide gas when it comes into contact with bleach or reagents containing bleach.
Data sourced from openFDA. This site is unofficial and independent of the FDA.