Recall Z-2949-2020— Intuitive Surgical, Inc.

Class II · Ongoing

Class II FDA medical device recall by Intuitive Surgical, Inc., initiated 2020-06-10. It names one FDA 510(k) clearance: K182188. Product: Ion Catheter Reprocessing Cover, IF1000 (Part Number:490114-02) UDI:00886874116845.. Reason: One type of the ION Catheter Reprocessing Covers (PN 490114-02) has the potential to become disengaged or incorrectly engaged to the ION Fully Articulating Catheter which can result in retention of cleaning chemical or water during reprocessing; may cause failure of the ION Fully Articulating Catheter or ION System..

Recalling firm
Intuitive Surgical, Inc.
Classification
Class II
Status
Ongoing
Initiated
2020-06-10
Firm location
Sunnyvale, CA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Ion Catheter Reprocessing Cover, IF1000 (Part Number:490114-02) UDI:00886874116845.

Reason for recall

One type of the ION Catheter Reprocessing Covers (PN 490114-02) has the potential to become disengaged or incorrectly engaged to the ION Fully Articulating Catheter which can result in retention of cleaning chemical or water during reprocessing; may cause failure of the ION Fully Articulating Catheter or ION System.

Data sourced from openFDA. This site is unofficial and independent of the FDA.