Recall Z-2929-2018— Siemens Healthcare Diagnostics, Inc.

Class II · Terminated

Class II FDA medical device recall by Siemens Healthcare Diagnostics, Inc., initiated 2018-07-16, terminated 2020-06-23. It names one FDA 510(k) clearance: K031560. Product: RAPIDLab 1245 Blood Gas Analyzer; 10321844, 10337179, 10491393 Device is point-of-care system for comprehensive lab-quality testing to patient's bedside. Test menu includes blood gas, electrolytes, glucose, lactate, neonatal total bilirubin and full CO-oximetry including total hemoglobin.. Reason: Therapeutic levels (1mg/mL and 2mg/mL) of Hydroxocobalamin may interfere with tHb and some of the CO-Ox fractions that are reported on the RAPIDPoint and RAPIDLab Blood Gas systems indicated. This may cause lower than expected values for carboxyhemoglobin (fCOHb) and methemoglobin (fMetHb). A negative interference with fCOHb has the potential to alter the medical assessment of the patient and may withhold necessary follow-up treatment and/or initiate cessation of medication in response to elevated fMetHb levels..

Recalling firm
Siemens Healthcare Diagnostics, Inc.
Classification
Class II
Status
Terminated
Initiated
2018-07-16
Terminated
2020-06-23
Firm location
Newark, DE, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

RAPIDLab 1245 Blood Gas Analyzer; 10321844, 10337179, 10491393 Device is point-of-care system for comprehensive lab-quality testing to patient's bedside. Test menu includes blood gas, electrolytes, glucose, lactate, neonatal total bilirubin and full CO-oximetry including total hemoglobin.

Reason for recall

Therapeutic levels (1mg/mL and 2mg/mL) of Hydroxocobalamin may interfere with tHb and some of the CO-Ox fractions that are reported on the RAPIDPoint and RAPIDLab Blood Gas systems indicated. This may cause lower than expected values for carboxyhemoglobin (fCOHb) and methemoglobin (fMetHb). A negative interference with fCOHb has the potential to alter the medical assessment of the patient and may withhold necessary follow-up treatment and/or initiate cessation of medication in response to elevated fMetHb levels.

Data sourced from openFDA. This site is unofficial and independent of the FDA.