Recall Z-2921-2020— Siemens Medical Solutions USA, Inc

Class II · Terminated

Class II FDA medical device recall by Siemens Medical Solutions USA, Inc, initiated 2020-07-24, terminated 2021-11-02. It names one FDA 510(k) clearance: K150493. Product: Siemens Sensis Vibe Hemo system in combination with the MicroPodTM EtCO2 module-diagnostic and administrative tool supporting hemodynamic catheterization and/or electrophysiology studies, for cardiac as well as interventional Radiology.. Reason: The connector linking the EtCO2 module to the HemoBox may break and become damaged, if the connector is re-attached to the HemoBox in this damaged state, under certain incorrect pin alignment conditions, the HemoBox may stop functioning and all vital signs are lost.

Recalling firm
Siemens Medical Solutions USA, Inc
Classification
Class II
Status
Terminated
Initiated
2020-07-24
Terminated
2021-11-02
Firm location
Malvern, PA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Siemens Sensis Vibe Hemo system in combination with the MicroPodTM EtCO2 module-diagnostic and administrative tool supporting hemodynamic catheterization and/or electrophysiology studies, for cardiac as well as interventional Radiology.

Reason for recall

The connector linking the EtCO2 module to the HemoBox may break and become damaged, if the connector is re-attached to the HemoBox in this damaged state, under certain incorrect pin alignment conditions, the HemoBox may stop functioning and all vital signs are lost

Data sourced from openFDA. This site is unofficial and independent of the FDA.