Recall Z-2911-2026— Contego Medical, Inc
Class II · Ongoing
Class II FDA medical device recall by Contego Medical, Inc, initiated 2026-06-29. FDA's recall record names no specific 510(k) clearance for this event. Product: Neuroguard IEP 3 in 1 Carotid Stent and Post-Dilation Balloon System with Integrated Embolic Protection (Neuroguard IEP System), REF: NG-0630-140-2, NG-0640-140-2, NG-0730-140-2, NG-0740-140-2, NG-NV-6-30, NG-NV-6-40, NG-NV-7-30, NG-NV-7-40. Reason: Carotid stent and post-dilation balloon system with integrated embolic protection system has an integrated filter, that due to a wire-to-filter-bond issue, may fail to open following sheath retraction; and if this condition were to occur during a procedure, the result may be loss of intended integrated embolic protection..
- Recalling firm
- Contego Medical, Inc
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2026-06-29
- Firm location
- Raleigh, NC, United States
Product description
Neuroguard IEP 3 in 1 Carotid Stent and Post-Dilation Balloon System with Integrated Embolic Protection (Neuroguard IEP System), REF: NG-0630-140-2, NG-0640-140-2, NG-0730-140-2, NG-0740-140-2, NG-NV-6-30, NG-NV-6-40, NG-NV-7-30, NG-NV-7-40
Reason for recall
Carotid stent and post-dilation balloon system with integrated embolic protection system has an integrated filter, that due to a wire-to-filter-bond issue, may fail to open following sheath retraction; and if this condition were to occur during a procedure, the result may be loss of intended integrated embolic protection.
Data sourced from openFDA. This site is unofficial and independent of the FDA.