Recall Z-2909-2020— CareFusion 303, Inc.
Class I · Ongoing
Class I FDA medical device recall by CareFusion 303, Inc., initiated 2020-08-04. It names one FDA 510(k) clearance: K133532. Product: Alaris System PC Unit Model 8015. Reason: The pump module keypad may exhibit keys that are unresponsive or stuck as a result of fluid ingress potentially resulting in a delay to the start of infusion, interruption of infusion, or inability to titrate medication. For high-risk patient populations who are receiving life-sustaining infusions, delays in an infusion or interruption can cause serious injury or death..
- Recalling firm
- CareFusion 303, Inc.
- Classification
- Class I
- Status
- Ongoing
- Initiated
- 2020-08-04
- Firm location
- San Diego, CA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Alaris System PC Unit Model 8015
Reason for recall
The pump module keypad may exhibit keys that are unresponsive or stuck as a result of fluid ingress potentially resulting in a delay to the start of infusion, interruption of infusion, or inability to titrate medication. For high-risk patient populations who are receiving life-sustaining infusions, delays in an infusion or interruption can cause serious injury or death.
Data sourced from openFDA. This site is unofficial and independent of the FDA.